Engineering, 28.06.2021 23:40 sjdswim

A pressure transducer has the following specifications: Input rage: 0-100 psi and the corresponding Output Range: 0-5 Volts. Linearity Error: 0.10% of the Reading Hysteresis Error: 0.10% of the Reading Sensitivity Error: 0.15% of the Reading Zero Drift Error: 0.20% of the Reading Its output is read via a voltmeter with instrument error of 0.10% of the reading and resolution of 0.01 V. If the applied pressure on the transducer is 65 psi, what is the design stage uncertainty of this pressure measurement system?

ansver
Answers: 3

Another question on Engineering

question
Engineering, 04.07.2019 18:10
What difference(s) did you notice using a pneumatic circuit over hydraulic circuit.explain why the pneumatic piston stumbles when it hits an obstacle.
Answers: 2
question
Engineering, 04.07.2019 18:10
Slip occurs via two partial dislocations because of (a) the shorter path of the partial dislocation lines; (b) the lower energy state through partial dislocations; (c) the charge balance.
Answers: 1
question
Engineering, 04.07.2019 18:10
Ifa component is made of two or more materials with different modulus of elasticity (e), it is called a composite member and we calculate the factor·n". mention the formula for calculating n". also, ifn> 1, explain what will happen to the 1. transformed.gi) ifn 1, what will happen to the material when transformed material when
Answers: 1
question
Engineering, 04.07.2019 18:20
Determine the damped natural frequencies and the steady state response of a decoupled damped forced two degrees of freedom system. 10ä1 + 2q1 20q1 10 cos t; 10q2 +4q2 + 40q2 10 cos t
Answers: 3
You know the right answer?
A pressure transducer has the following specifications: Input rage: 0-100 psi and the corresponding...